Siemens ACUSON SC2000 PRIME

Siemens Healthineers · ACUSON series

Cart-basedCardiology

Siemens ACUSON SC2000 PRIME cart-based ultrasound system

The Siemens ACUSON SC2000 PRIME is a cardiovascular ultrasound system for 2D and 4D transthoracic, transesophageal and intracardiac echocardiography. It is designed for structural heart disease work, with TrueFusion echo-fluoro guidance and AI-powered eSie measurement packages. The PRIME edition added the Z6Ms 4D TEE transducer and eSie Valves to the SC2000, both offered as upgrades for earlier systems.

Released
2014launched by Siemens
Probe ports
3universal
Display
21.5-inch54.5 cm diagonal, 16:9
Weight
136 kg299 lbs

Compatible probes

Linear

  • 9L4

Convex & microconvex

  • 6C1 HD

Specialty

  • V5Ms 2D transesophageal
  • V7M 2D transesophageal, adult and pediatric
  • Z6Ms 4D transesophageal, introduced with the PRIME
  • Aux CW Pencil CW
  • CW2 Pencil CW
  • ACUSON AcuNav 8F ICE Catheter 2D intracardiac
  • ACUSON AcuNav 10F ICE Catheter 2D intracardiac
  • ACUSON AcuNav Volume ICE Catheter 4D intracardiac, Release 5.1 and later
  • ACUSON AcuNav V Ultrasound Catheter Earlier 4D intracardiac catheter
  • AcuNav Crystal 2D ICE Catheter 2D intracardiac, on v6.5 (VC15)
  • SOUNDSTAR 10F ICE Catheter Third-party intracardiac catheter
  • SOUNDSTAR eco 10F ICE Catheter Third-party intracardiac catheter
  • SOUNDSTAR eco 8F ICE Catheter Third-party intracardiac catheter
  • SoundStar 8F Third-party intracardiac catheter, on VC11 (v6.1)
  • SoundStar Crystal Ultrasound Catheter Catheter, on v6.5 (VC15)

Specifications

Display
Monitor
54.5 cm (21.5 in) diagonal, 16:9 widescreen format
Physical
Weight
136 kg (299 lbs)
Dimensions
Width 64 cm (25.2 in), height 109 cm (42.9 in), depth 177–183 cm (69.7–72 in)
Probe ports
Three universal transducer ports
Controls
Control panel lock, table-side remote-control joystick, rear handle, swivel castersShow all
Connectivity & options
DICOM
Storage, storage commitment, query/retrieve, modality worklist, MPPS, structured reportsShow all
Network protocol
IPv4 supported, IPv6 not supported (VC15)
Wireless
Wireless connectivity with WiFi EAP-TLS authentication, on VC11 (v6.1)
Remote service
Kinectus Remote Service, SRS (smart remote services), eSieLink Remote Assistance
Offline analysis
syngo SC2000 Workplace, syngo Ultrasound Apps Suite, syngo Dynamics, TOMTEC-ARENA
SwiftLink Plus Connector
Port-adapter cable for AcuNav Crystal and SoundStar Crystal catheters, on v6.5
Imaging & software
Imaging modes
2D, THI, M-mode, color M-mode, color Doppler, PW/CW Doppler, bi-plane, real-time 4DShow all
Contrast
LVO contrast imaging; Contrast Pulse Sequencing (CPS) for TTE
Echocardiography
2D and 4D TTE, TEE and ICE, with TrueFusion
Imaging technologies
InFocus Coherent Technology, True Fidelity Color Doppler, 64-beam technology
Real-time Volume Imaging
Without gating or stitching
Real-Time Volume Color Doppler
Non-stitched volume color Doppler at 21 volumes per second
Volume tools
eSie Lines, eSie Slice, Single V, Dual V with real-time volume color Doppler
Image processing
On v6.1: Advanced SieClear spatial compounding, Clarify vascular enhancement, TEQShow all
2D applications
eSie Measure (B-mode, M-mode, spectral Doppler), eSie Left Heart, eSie VVIShow all
4D applications
eSie LVA, eSie PISA, eSie Valves, RVA, TrueFusion; availability varies by transducerShow all
Stress echo
Rapid Stress Volume Stress Echo App., on v6.1
eSie Flow
Limited-release offering for qualified expert healthcare professionals
Interventional guidance
Septal Guide and Septal Guidelines for septal puncture; TrueFusion echo-fluoro guidanceShow all
ICE Probe Saver
On Release 6.5, lowers transmit power automatically when it detects ICE imaging in air
Revision and software
Release 5.0, 5.1, 6.0 and 6.5; software v6.0 = VC10, v6.1 = VC11, v6.5 = VC15Show all
Operating system
Microsoft Windows 10, on VC11 (v6.1)
Compliance
FDA 510(k)
K233613 (2024-06-12), K211726, K202683, K170315, K162221, K142628 (2014-10-10)Show all

What to check when buying a used Siemens ACUSON SC2000 PRIME

  1. Test every TEE probe for leakage and wear

    A TEE probe with cuts, tears, fraying or loose parts must not be used, and V5Ms probes were recalled for deterioration of the material covering the articulating section.

    How: Get a leakage test result for each TEE probe offered, and inspect the insertion shaft, distal tip and articulating section for cuts, tears, fraying, wear or loose parts.

  2. Check each Z6Ms sleeve, serial number and leakage

    Recall Z-0667-2017 covered Z6Ms probes for acquisition errors and deterioration of the material covering the articulating section, and a later report traced a failed leakage test to a hole in the articulation sleeve.

    How: Compare the probe's serial number with the production changes that moved the sleeve from C-Flex to Viton and addressed Gastro-flex circuit corrosion, then check the sleeve for holes or wear and get a leakage test.

  3. Confirm the MPI board fix before using V5M

    With MPI board version 4, recall Z-1081-2013 found the V5M transducer's displayed temperature could read lower than its actual temperature, with a risk of esophageal burns.

    How: Read the MPI board revision and ask for the service record showing Siemens exchanged the affected board.

  4. Check the user interface module is firmly attached

    Recall Z-0824-2014 found the user interface assembly could work loose, and the whole module could fall off the system.

    How: On units in the recall's serial range, look for any looseness where the module attaches to the column, and stop using a system that shows it.

Common questions about the Siemens ACUSON SC2000 PRIME

Is the Siemens ACUSON SC2000 PRIME FDA-cleared?

Yes. It is FDA-cleared under 510(k), and its clearances use the device name ACUSON SC2000 Diagnostic Ultrasound System rather than the PRIME name. K142628, decided October 10, 2014, added the Z6Ms transducer that came with the PRIME edition to the SC2000 cleared under K132654.

Later clearances include K211726, decided November 5, 2021, for software VC11 (v6.1) with syngo TrueFusion 2.0, and K233613, decided June 12, 2024, for AcuNav Crystal and SoundStar Crystal catheter compatibility on v6.5 (VC15).

Has Siemens ended support for the ACUSON SC2000 PRIME?

Not by name. The ACUSON SC2000 is on Siemens Healthineers' US end-of-support list, with an upgrade to the ACUSON Origin, but that entry does not name the PRIME and gives no date.

Is the AcuNav Volume ICE catheter only for the ACUSON SC2000 PRIME?

Yes. Siemens Healthineers offers the ACUSON AcuNav Volume ICE Catheter, a 4D intracardiac catheter, exclusively on the ACUSON SC2000 PRIME, and the system needs Release 5.1 or later to run it.