Siemens ACUSON NX3 Elite

Siemens Healthineers · ACUSON series

Cart-basedCardiologyObstetrics & GynecologyVascular

Siemens ACUSON NX3 Elite cart-based ultrasound system

The Siemens ACUSON NX3 Elite is a mid-range, cart-based ultrasound system mainly for general imaging, shared service cardiology and OB/Gyn. It shares the NX3 Series with the ACUSON NX3 and adds cleared intra-operative and cardiac indications, including transesophageal, plus eSie Touch and contrast agent imaging. It weighs 75 kg without peripherals, with a 21.5-inch display and four active probe ports.

Released
2015launched by Siemens
Probe ports
4active
Display
21.5-inch1920 x 1080 LED, IPS
Weight
75 kg165 lb, without peripherals

Compatible probes

Linear

  • 11L4
  • VF10-5
  • VF12-4
  • VF13-5SP Intraoperative
  • VF16-5

Convex & microconvex

  • CH5-2
  • C8-5
  • C8F3 4D transducer

Phased

  • P4-2
  • P8-4

Endocavity

  • EC9-4
  • EC10-5w 220° field of view
  • BP10-3 Biplane, for prostate
  • 10MC3

Specialty

  • V5Ms Transesophageal
  • CW2 Non-imaging continuous wave Doppler
  • CW5 Non-imaging continuous wave Doppler

Specifications

Display
Display
21.5-inch widescreen color LED flat panel, IPS, 1920 x 1080, 178° angle of viewShow all
Touch screen
10.4-inch touch screen on the control panel, 1024 x 768; supports swipe interaction
Monitor arm
Articulating: arm and panel rotation -80° to +80°, tilt 15° to 70°, pull 210 mmShow all
Physical
Weight
75 kg (165 lb) without OEM peripherals
Dimensions
Height 144.7–158.6 cm (57.0–62.4 in); width 54 cm (21.3 in); depth 92.3 cm (36.3 in)
Wheels
Four 125 mm casters, 2 front and 2 rear, each with total lock
Probe ports
4 active ports for phased, curved and linear array transducers; DL (260) connectorsShow all
Control panel
Home Base design, height-adjustable 11 cm; backlit keys, soft keys, QWERTY keyboardShow all
Transducer holders
5 configurable holders with gel bottle storage, plus a dedicated endocavity holderShow all
Cooling
Air cooling
Power & battery
Power
100-120/200-240 VAC, 50/60 Hz; isolation transformer; max 600 VA with OEM peripheralsShow all
QuikStart standby (option)
Power-down in 10 sec; complete boot-up in less than 10 sec; standby time > 35 minShow all
Start-up and shut-down (NX3 Series)
1 minute 47 seconds from off-mode to operating state; 20 seconds back to off-mode
Energy use (NX3 Series)
0.23 kW average during examinations; 0.13 kW operating without an examination
Connectivity & options
DICOM
DICOM 3.0: store, print, storage commitment, worklist, MPPS, media export; SR (option)Show all
Network and wireless
Ethernet RJ45 (10/100/1000 BaseT); wireless option via USB dongle, 802.11 a/b/g/n/acShow all
Ports
6 USB 2.0 (2 front, 4 rear); VGA out, DVI out, DVI in/out; audio out; 2 AC outletsShow all
Storage and media
500 GB drive, 300 GB for patient data; up to 300,000 images with compressionShow all
Documentation devices
Up to 2 on-board devices in the cart: a B/W printer and a color printer or DVR
Physiological interface
3-lead ECG, auxiliary physio input, R-wave single and dual trigger, respiratory traceShow all
Workflow and hardware options
QuikStart, gel warmer, wireless transfer, barcode reader, DICOM SR, dual foot switchShow all
Remote service
eSieLink Remote Assistance; Smart Remote Service and eSiePrep setup (options)Show all
System security
Embedded virus protection
Imaging & software
System architecture
All-digital beamformer, parallel Quad beam processing; up to 172,032 processing channelsShow all
Frequency range
1.3–16.0 MHz; MultiHertz: up to seven 2D and THI and up to two Doppler frequenciesShow all
Image depth
1–30 cm in 1.0 cm increments (transducer dependent)
Imaging modes
2D with THI; color, power and directional power Doppler; PW, CW and steerable CW; M-modeShow all
Imaging technologies
DTO, Dynamic TCE, TGO, HD Zoom, Advanced SieClear, SieClear, SieScape, Clarify VEShow all
Imaging options
DTI, eSie Touch elasticity, Cadence contrast agent imaging, anatomical M-modeShow all
Measurement and analysis options
syngo Auto OB, Auto Follicle, Auto Left Heart, AHP, Axius EF, Stress Echo, PhysioShow all
3D/4D (option)
3-Scape freehand 3D; fourSight 4D on the C8F3, up to 30 vol/sec; Advanced fourSightShow all
Cardiac and TEE (2018 configuration)
fourSight TEE imaging, syngo VVI and syngo Mitral Valve Assessments (MVA)
Revision and software
VA10 (August 2015) and VB20 (December 2017, software ID 5.1.xxx)Show all
Environment
Operating environment
10–40 °C (50–104 °F) without OEM peripherals; 30–80% humidity non-condensingShow all
Heat output and noise
Maximum heat output 2150 BTU/hr; average noise level 46–47 dB
Compliance
FDA 510(k)
K152469 (2015-09-21), K173957 (2018-02-01), K192835 (2019-10-22), K202683 (2020-10-15)Show all
Standards
ISO 9001, ISO 13485, 21 CFR 820; UL 60601-1, EN/IEC 60601-1, IEC 60601-2-37Show all
Languages
Languages
Screen text and panel overlay: Chinese, English, French, Italian, German, SpanishShow all

What to check when buying a used Siemens ACUSON NX3 Elite

  1. Confirm the acoustic output recall correction

    FDA recall Z-2353-2019 found that transducers on affected software could exceed their acoustic output power values in pulsed wave Doppler, and the recall is listed as open, with not every unit corrected.

    How: Read the software version: VA10A to VA10F, VB20A and VB20B are listed as affected, and VB20C is the corrected VB20 release.

  2. Look for fluid ingress at panel seams

    Fluid that seeps into the electrical circuitry can cause excessive leakage current or system failure, and spray cleaners can force it in.

    How: Inspect the control panel seams, the keyboard and the DVD slot for any sign that fluid has seeped inside.

  3. Inspect the power cable for heat damage

    An FDA report describes an NX3 Series system whose power cable melted at a wall outlet with old wiring.

    How: Look along the cable and at the plug for melted or heat-marked insulation, and check the outlet the system will use.

  4. Have the V5Ms TEE probe leakage-tested

    The V5Ms needs an electrical leakage current test before each use, and one that fails again after its connections are checked must not be used on a patient.

    How: Ask for a leakage test on the V5Ms, or run one, and do not accept a probe that fails it.

Common questions about the Siemens ACUSON NX3 Elite

Is the Siemens ACUSON NX3 Elite FDA-cleared?

Yes. It is FDA-cleared under 510(k) K152469, decided September 21, 2015, as the ACUSON NX3 Elite / NX3 Diagnostic Ultrasound System, a Class II device under 21 CFR 892.1550 with product codes IYN, IYO and ITX.

Later clearances cover it too: K173957, decided February 1, 2018; K192835, decided October 22, 2019, which extended its cardiac indication to include transesophageal use; and K202683, decided October 15, 2020, which added lung-ultrasound imaging to the labeling with no new features or software changes.

What is the difference between the ACUSON NX3 Elite and the ACUSON NX3?

Both belong to Siemens' NX3 Series and are cleared under the same 510(k)s. The NX3 Elite's cleared indications add intra-operative use (abdominal, small parts and vascular) and cardiac use including transesophageal, which the NX3's indications do not list.

The NX3 Elite also offers syngo Arterial Health Package, syngo Auto Left Heart, contrast agent imaging and eSie Touch elasticity imaging, features the clearances list for the NX3 Elite only.

Can the Siemens ACUSON NX3 Elite do transesophageal echo?

Yes, with the V5Ms TEE transducer. K192835, decided October 22, 2019, extended the NX3 Elite's cardiac indication to include transesophageal use, and fourSight TEE imaging is listed for the NX3 Elite configuration cleared under K173957 on February 1, 2018.

Is the Siemens ACUSON NX3 Elite discontinued?

Yes. Its FDA device record gives December 31, 2025 as the end of its commercial distribution. Siemens launched it alongside the ACUSON NX3 at Medica 2015 in Düsseldorf.

Where is the Siemens ACUSON NX3 Elite made?

Siemens Healthineers builds it at its facility in Seongnam-si, Gyeonggi-do, in the Republic of Korea.