Siemens ACUSON NX3

Siemens Healthineers · ACUSON series

Cart-basedCardiologyObstetrics & GynecologyVascular

Siemens ACUSON NX3 cart-based ultrasound system

The Siemens ACUSON NX3 is a mid-range cart-based ultrasound system for general imaging, shared service cardiology and OB/GYN in hospitals and clinics. It weighs 75 kg (165 lb), with a 21.5-inch LED display and up to 4 active transducer ports. Its NX Series siblings include the NX2, with a deliberately limited probe selection, and the NX3 Elite, which adds intra-operative and transesophageal indications.

Released
2015first FDA clearance
Probe ports
Up to 4active transducer ports
Display
21.5-inchLED; 10.4-inch touch screen
Weight
75 kg165 lb, without OEMs

Compatible probes

Linear

  • VF12-4
  • 11L4 · 3.5–12.8 MHz For NX3 R2.0
  • VF10-5

Convex & microconvex

  • CH5-2 · 1.4–5 MHz
  • C8F3 4D

Phased

  • P4-2

Endocavity

  • EC9-4 · 3.1–10 MHz
  • 10MC3 · 3.1–10 MHz For NX3 R2.0
  • BP10-3 · 3.3–9.9 MHz Bi-plane prostate imaging; for NX3 R2.0

Specifications

Display
Display
21.5-inch high-resolution flat-panel LED widescreen color display
Touch screen
10.4-inch high-resolution touch display
Physical
Weight
75 kg (165 lb) without OEMs
Dimensions
Width 54 cm (21.3 in), height 149.7 cm (58.9 in), depth 92.3 cm (36.3 in)
Probe ports
Up to 4 active transducer ports
Controls
Up to 28% fewer keystrokes and 3x more user-customizable settings (data on file)Show all
Packaging
Approx. 35 kg (77.2 lb): wood, paper/cardboard and plastic
Power & battery
Power consumption
0.21 kW ramp-up, 0.13 kW standby, 0.23 kW in scan mode
Power-on and power-off
Power-on 1 min 47 s; power-off 20 s; complete switch-off possible
Connectivity & options
DICOM
DICOM 3.0: Storage, Storage Commitment, Query/Retrieve, Modality Worklist, MPPS, PrintShow all
Removable media
DICOM on 120 mm CD-R, CD-RW, DVD-R, DVD+R, DVD-RW, DVD+RW
Network
Wired IEEE 802.3 10BaseT/100BaseT/1000BaseT; wireless
Remote service
SRS (Smart Remote Service), optional 24/7, with preinstalled internet-based connectivity
Remote assistance
eSieLink: secure, real-time remote support session from the system
Pre-shipment setup
eSiePrep: internet-based connectivity configured before shipping
Imaging & software
Imaging modes
B-mode, M-mode, PW, steerable CW, color, power and directional power Doppler, THIShow all
Image processing
Clarify VE, DTCE, SieClear and Advanced SieClear, TGO, MultiHertz, DIMAQ, HD ZoomShow all
3D, 4D and panoramic
3-Scape 3D with 3D measurements; syngo fourSight 4D (basic); SieScape panoramic imaging
Cardiac
Stress Echo, Axius EF, anatomical M-mode (AMM), Doppler tissue image, syngo VVIShow all
OB/GYN
syngo Auto OB measurements (option); syngo Auto follicle
Vascular
One-touch color Doppler flow states for vascular exams
Bi-plane imaging
Live simultaneous bi-plane imaging with the BP10-3, side-by-side or top/bottom split
Revision and software
Releases VA10 and VB20; builds VA10A–VA10F, VB20A and VB20BShow all
Operating system
Microsoft Windows 7-based digital architecture
Environment
Operating environment
10–40 °C room temperature; 10–80% relative humidity
Noise level
46.9 dB(A) at basic load (without probe) and at full load (with the VF10-5 probe)Show all
Heat emission
0.13 kW in standby, 0.23 kW in scan mode; air-cooled, no heat-recovery interface
Compliance
FDA 510(k)
K152469 (2015-09-21), K173957 (2018-02-01), K192835 (2019-10-22), K202683 (2020-10-15)Show all
Model number
NX3 11235612 (UDI 04056869004297); NX3 Elite 11235411
Manufacturing site
Seongnam-si, Gyeonggi-do, Republic of Korea

What to check when buying a used Siemens ACUSON NX3

  1. Check the software against recall Z-2352-2019

    On NX3 units running software VA10A to VA10F, VB20A or VB20B, transducers could exceed their defined acoustic output in PW Doppler because transmit voltages were not loaded correctly.

    How: Read the installed software version and ask for the service record showing Siemens' corrective software update.

  2. Inspect each probe's housing, lens and cable

    A damaged probe must not be used.

    How: Examine the housing, acoustic lens, strain relief and cable of every probe offered, and set aside any probe showing cracks, splits or other damage.

  3. Confirm the model label and each probe's clearance

    The NX3 and the NX3 Elite are separate models, and transducers such as the EC10-5w, VF16-5 and V5Ms are FDA-cleared for the NX3 Elite only.

    How: Read the model number on the system label, and confirm that every transducer offered is one cleared for the NX3: the CH5-2, VF10-5, VF12-4, EC9-4, P4-2, C8F3, 11L4, 10MC3 or BP10-3.

  4. Restart it several times, including from standby

    One NX3 Series system would not power on after a reboot, a fault attributed to an unstable cable connection or improper standby use.

    How: Run several cold starts and standby-and-resume cycles, then reseat and inspect the cable connections.

Common questions about the Siemens ACUSON NX3

Is the Siemens ACUSON NX3 FDA-cleared?

Yes. The ACUSON NX3 was first FDA-cleared under 510(k) K152469, decided on September 21, 2015, under regulation 21 CFR 892.1550 with product codes IYN, ITX and IYO. Later clearances are K173957 (February 1, 2018), K192835 (October 22, 2019) and K202683 (October 15, 2020, a multi-system ACUSON submission), and each names both the NX3 and the NX3 Elite.

Has the Siemens ACUSON NX3 been recalled?

Yes. FDA recall Z-2352-2019, initiated by Siemens on January 28, 2019, covers NX3 systems running software VA10A through VA10F, VB20A or VB20B, on which PW Doppler could exceed the defined acoustic output power. Its cause was software design, and in September 2026 its FDA status was Open, Classified.

What is the difference between the Siemens ACUSON NX3 and NX3 Elite?

Both belong to the ACUSON NX3 Series and have a 21.5-inch flat-panel display. In their FDA clearances, the NX3 Elite adds intra-operative indications, transesophageal cardiac use, 12.0 and 13.0 MHz center frequencies, continuous-wave Doppler (CWD), Auto Left Heart and syngo AHP (Arterial Health Package).

Some transducers are cleared for the NX3 Elite only: the EC10-5w, CW2, CW5, VF16-5, VF13-5sp, C8-5, P8-4 and V5Ms. The model number tells the two apart: 11235612 is the NX3 and 11235411 the NX3 Elite.

What exams is the Siemens ACUSON NX3 cleared for?

Its U.S. indications for use under K192835 are fetal, abdominal (including liver), pediatric, small parts (small organ), neonatal cephalic, adult cephalic, transcranial, OB/GYN, pelvic, neonatal, cardiac, vascular (including peripheral vessel), musculoskeletal, superficial musculoskeletal and urology imaging.