Siemens ACUSON NX2
Siemens Healthineers · ACUSON series

The Siemens ACUSON NX2 is a cart-based general imaging system in the ACUSON NX Series. Its FDA-cleared applications include cardiac, OB/GYN, vascular and musculoskeletal imaging, and the C5-2v and L10-5v transducers were introduced with it. The ACUSON NX2 Elite adds the CW2 and CW5 transducers, Advanced fourSight 4D and eSie Touch elasticity imaging.
- Released
- 2016first FDA clearance
- Probe ports
- 3 active260-pin; 4th port optional
- Display
- 21.5 in1080p LCD, articulating arm
- Weight
- 62 kg137 lb, without peripherals
Compatible probes
Linear
- L10-5v Introduced with the NX2
- VF10-5
Convex & microconvex
- C5-2v Introduced with the NX2
- CH5-2
- C8F3 For the fourSight 4D option
Phased
- P4-2
Endocavity
- EC9-4
- 10MC3 Cleared in K173981
Specifications
Display
- Monitor
21.5-inch LED LCD, 1920 x 1080, IPS, widescreen, on an articulating arm21.5-inch flat panel display monitor: active-matrix LCD, 1920 x 1080 resolution (1080p HD); LED, widescreen, color, progressive scan (non-interlaced), IPS (in-plane switching); brightness control; mounted on an articulating arm.Show allShow less
- Tilt and swivel
- Tilt 15° up and 80° down; swivel 160°
Physical
- Weight
- 62 kg (137 lb) without documentation devices; 92 kg with safety working load
- Dimensions (maximum)
- Width 532 mm (20.9 in); height 1431 mm (56.3 in); depth 862 mm (33.9 in)
- Wheels and brakes
- Four locking brakes; each wheel has a locking lever and a release lever
- Control panel
Backlit, alphanumeric keyboard, trackball, soft keys, task lightingBacklit control panel with an alphanumeric keyboard and special function keys; task lighting; language-specific control panel overlays; trackball with SET key; soft-key selections at the bottom of the screen, worked with toggle keys and a PAGE control; user-programmable keys.Show allShow less
- Height adjustment
- Adjustable in height for the user
Power & battery
- Power supply
100–120 V~ (90–132 V~) or 200–240 V~ (180–264 V~), 50/60 Hz100–120 V~ (range 90–132 V~), 4.0–3.5 A maximum, 50/60 Hz; or 200–240 V~ (range 180–264 V~), 2.5–2.0 A maximum, 50/60 Hz.Show allShow less
- Maximum power consumption
- 600 VA with OEM peripherals
- Power consumption by mode
0.21 kW scanning; 0.20 kW ready to scan; 0.14 kW stand-by0.19 kW during ramp-up; 0.20 kW on, ready to scan; 0.21 kW scanning; 0.20 kW freeze mode; 0.20 kW screen saver mode; 0.14 kW stand-by.Show allShow less
- Battery
- Installed battery pack for QuikStart battery-powered standby (QuikStart option)
- Power-on time
1 min 59.92 s (virus protection on); 1 min 42.98 s (off); power-off 22.28 sOff to ready to scan: 1 minute 59.92 seconds with virus protection on, 1 minute 42.98 seconds with virus protection off. Power-off: 22.28 seconds.Show allShow less
Connectivity & options
- DICOM
DICOM 3.0 print and storage with worklist; MPPS and structured reporting optionsDICOM 3.0 Connectivity Option (print and storage, includes DICOM Worklist); DICOM Modality Worklist Option; DICOM MPPS Option; DICOM Structured Reporting for Obstetrics/Gynecology, Cardiac and Vascular. Application entities: Verification, Storage, Storage Commitment, Query/Retrieve, Worklist, MPPS, Print.Show allShow less
- Media
- Read-write CD/DVD combination drive; USB drive (DICOMDIR)
- USB
- Series A: 4 on the I/O panel, 2 on the control panel, 1 on the BW printer panel
- Network
- Ethernet 10BaseT/100BaseT, RJ45
- Video
HDMI 19-pin (1 in, 1 out); DVI 24-pin out; VGA 15-pin out; S-Video 4-pinHDMI listed as 19-pin DVI-type (1 input, 1 output); DVI 24-pin (1 output); VGA 15-pin D-sub (1 output); S-Video 4-pin.Show allShow less
- Audio
- 2-channel RCA, 2 in and 2 out
- Physio panel
- ECG connector and Aux 1 connector; ECG through the Physio Module option
- Recording devices
- On-board location for up to two image recording devices
- Wireless network (option)
802.11a/ac/b/g/n, 2.4 and 5 GHz; WPA, WPA2 (and PSK), open, shared; WEP, TKIP, AES802.11a/ac/b/g/n (and mixed); 2.4 GHz and 5 GHz; authentication WPA, WPA2, WPA-PSK, WPA2-PSK, open, shared; encryption WEP, TKIP, AES.Show allShow less
- Footswitch
- Optional user-programmable dual-pedal footswitch (USB)
- Remote service (SRS)
- Encrypted system monitoring, log retrieval and remote troubleshooting sessions
- eSieLink Remote Assistance
- Virtual communication option with headset and camera
- eSiePrep
- Sales option that configures internet connectivity for SRS before shipment
Imaging & software
- Revision and software
VA10 (software ID 5.1.xxx); VA11 (Product Version 1.1)VA10 (software ID 5.1.xxx) and VA11 (Product Version 1.1). Versions named in FDA recall Z-2350-2019: VA10A, VA10B, VA10C, VA11A, VA11B and VA11C.Show allShow less
- Imaging modes
2D, Dual, Split (fundamental, harmonic); M-mode; color, power, PW Doppler; 4B; AMMSingle modes: 2D-mode, Dual mode and Split mode in fundamental and harmonics; M-mode; Color Doppler; Power Doppler; Pulsed Wave Doppler. Mixed modes: 2D/M-mode; 2D/Doppler; 2D-mode with color (or power); Dual-mode and Split-mode with color (or power); 2D-mode with color/Doppler and with power/Doppler (update and triplex); 4B with color (or power); M-mode with color. Also 4B-mode and Anatomical M-mode (for cardiac imaging).Show allShow less
- Imaging technologies
Tissue Harmonic Imaging; Advanced SieClear compounding; Dynamic TCE; SieClear; TGOTissue Harmonic Imaging (THI); Advanced SieClear spatial compounding (ASC); Dynamic TCE tissue contrast enhancement (DTCE); SieClear; Tissue Grayscale Optimization (TGO); directional power Doppler imaging; split-screen 2D/Color display formats.Show allShow less
- Frequency range
- 2.0 to 10.0 MHz; Wideband MultiHertz multiple-frequency transducer technology
- 2D-mode
Depth 10 mm to 300 mm (transducer-dependent); up to 4 focal zones; zoom up to 10xDepth 10 mm to 300 mm (transducer-dependent); up to 4 focal zones; CINE 256 gray scale frames (128 frames for Dual mode); zoom up to 10x; transmit power 0 dB to -54 dB.Show allShow less
- 3D and 4D
- 3-Scape imaging option; fourSight 4D imaging option (includes 3-Scape; C8F3 transducer)
- Imaging options
- Clarify VE vascular enhancement; SieScape panoramic imaging; HD Zoom; Anatomical M-Mode
- Measurement packages
syngo Arterial Health (CIMT); syngo Auto Left Heart (P4-2); syngo Auto OBsyngo Arterial Health Package (carotid intima-media thickness, linear transducers); syngo Auto Left Heart (P4-2, cardiac); syngo Auto OB measurements.Show allShow less
- QuickSet presets
- Up to 128 user-programmable sets of system parameters
- Options limited to the NX2 Elite
- Advanced fourSight 4D and eSie Touch Elasticity Imaging
- Operating system
Windows-based; anti-virus by whitelisting; firewall with defined open portsWindows-based operating software; all-digital system architecture; anti-virus protection by whitelisting; firewall with defined open ports (DICOM on TCP 104).Show allShow less
Environment
- Operating conditions
- +10 °C to +40 °C; up to 3,000 m (700–1060 hPa)
- Storage and transport
- -20 °C to +55 °C; < 95% relative humidity, non-condensing; up to 5,000 m (500–1060 hPa)
- Heat emission
- < 0.20 kW ready to scan; < 0.21 kW scanning
- Noise
- 38.5 dB(A) ready to scan; 41.7 dB(A) scanning
- Cooling
- Air
Compliance
- FDA 510(k)
K161787 (July 28, 2016); K172374 (August 15, 2017); K173981 (February 2, 2018)K161787 (July 28, 2016), ACUSON NX2 Diagnostic Ultrasound Systems, product code IYN; K172374 (August 15, 2017), with the ACUSON NX2 Elite; K173981 (February 2, 2018), with the ACUSON NX2 Elite.Show allShow less
- Electric shock protection
- Class I, externally powered
What to check when buying a used Siemens ACUSON NX2
Confirm the PW Doppler recall correction
FDA recall Z-2350-2019 found that, on affected software, transducers could exceed their acoustic output values in PW Doppler because transmit voltage values were not loaded correctly into the hardware.
How: On units with VA10A, VA10B, VA11A or VA11B software, ask for the service record of the Siemens update, since a VA10C or VA11C version alone does not prove the correction.
Run an electrical safety and leakage test
Electrical safety testing is required for this system, and an extension cord or multi-socket outlet can compromise grounding and push leakage current over its limits.
How: Have the leakage test run with the system plugged straight into a hospital-grade outlet, as US use requires, and reject a unit that fails it.
Inspect the fixed power cord for wear
The ACUSON NX2's power cord is not detachable, so a frayed, split or worn cord means a Siemens service replacement.
How: Do not power on a unit whose cord is frayed, split, worn or tampered with, or whose ground plug is bypassed with an adapter.
Look for fluid inside the console
Fluid that seeps into the electrical circuitry can cause excessive leakage current or system failure, and liquid can damage the control panel.
How: Look for signs that liquid has entered the console or control panel, and follow up any electrical or functional fault found alongside them.
Common questions about the Siemens ACUSON NX2
Is the Siemens ACUSON NX2 FDA-cleared?
Yes. The ACUSON NX2 was cleared under 510(k) K161787, with a decision date of July 28, 2016.
Two later 510(k)s cleared it together with the ACUSON NX2 Elite: K172374 on August 15, 2017, which treats the NX2 as the same system as in K161787, and K173981 on February 2, 2018, which lists the 10MC3 transducer for both systems.
What exams is the Siemens ACUSON NX2 cleared for?
Its first clearance, K161787, includes cardiac, fetal, abdominal (including liver), pediatric, small parts (small organ), transcranial, OB/GYN, pelvic, urology, vascular (including peripheral vessel) and musculoskeletal applications. K173981 adds adult cephalic.
What is the difference between the ACUSON NX2 and the NX2 Elite?
Both are ACUSON NX2 Series systems. The NX2 Elite adds the CW2 and CW5 pencil transducers, which are cleared for the Elite only, and the Advanced fourSight 4D and eSie Touch Elasticity Imaging features.
A fourth active transducer port and 3-Scape imaging are standard on the Elite and options on the NX2. The NX2 offers the fourSight 4D imaging option, which includes 3-Scape and uses the C8F3 transducer.
How is the ACUSON NX2 related to the ACUSON NX3?
The NX2 Series takes its industrial design from the ACUSON NX3. The NX2's first 510(k), K161787, named the ACUSON NX3 (K152469) and the ACUSON X700 (K141846) as predicate devices.