Mindray TE Air

Mindray · TE-Series

HandheldPoint of CareCardiology

Mindray TE Air handheld ultrasound system

The Mindray TE Air is a wireless handheld point-of-care ultrasound transducer, used with the TE Air Mobile App on a mobile device or with the TE X Ultrasound System. It is for scanning at the bedside, in emergency rooms and outside the hospital, and is FDA-cleared for cardiac, abdominal, thoracic and urology exams. Its i3P probe is marketed as the TE Air i3P; the TE Air-Series also includes the TE Air e5M.

Released
2023launched by Mindray
Probe ports
Wireless5 GHz Wi-Fi, no host cable
Display
None built inuses the host's screen
Weight
199 ± 3 gprobe end

Compatible probes

Phased

  • i3P · 2–4 MHz 64 elements; marketed as the TE Air i3P
  • i3PA

Specialty

  • i5M Convex and linear acoustic data
  • e5M Convex and linear; marketed as the TE Air e5M

Specifications

Display
Display
None built in; the TE Air app uses the host device's screenShow all
Minimum host display
iOS 4.7 in, 1334 × 750; Android phone 6.41 in, 2340 × 1080; pad 10.5 in, 2560 × 1600Show all
Display brightness
Host minimum 400 nit for iOS and Android phones, 200 nit for the Android padShow all
Physical
Weight
199 ± 3 g for i3P and i3PA; up to 229 ± 3 g for other probe modelsShow all
Dimensions
46.5 × 33 × 170 mm for i3P and i3PA; 70 × 33 × 170 mm for i5M, z5M, e5M and a5Mx modelsShow all
Controls
Touch control in the app, designed for one hand; probe power key and multifunction buttonShow all
Water and dust rating
IP68 waterproof and dustproof; immersible in compatible disinfectantsShow all
Power & battery
Battery
Built-in probe battery, 3.85 V DC, 1650 mAh
Scanning time
Over 60 min of scanning; at least 80 min in B mode at 40% freeze, 60% scan timeShow all
Charging
Magnetic charging cable, 5 V or 9 V input, 3 A max; 0–100% in no more than 45 minShow all
Air Capsule
Case and charger in one; up to three wireless recharges; 3.6 V DC, 3000 mAh batteryShow all
Connectivity & options
Probe-to-host connection
Wi-Fi, IEEE 802.11 a/b/g/n/ac at 5 GHz, WPA2 (PSK) encryption; no cable to the host
Host devices
TE Air app on iOS 13.0+ or Android 10.0+ mobile devices, or the TE X Ultrasound SystemShow all
Minimum host hardware
iOS: A10, 128 GB, 2 GB RAM; Android phone: Snapdragon 855, 256 GB, 8 GB RAM…Show all
Connectivity
PACS; DICOM/HL7 over Wi-Fi; DICOM storage and WorkList with DICOM modules; LDAP, TricefyShow all
Data security
Role-based access control; patient data encrypted locally and in transmission
Standard configuration
Probe with charging cable, system software, accessories and operator's manuals
Options
Air Capsule, TDI, Extended Connection, AutoEF, iNeedle, iTouch+, Smart Calc, iWorks…Show all
Imaging & software
Imaging modes
B, M, Color Doppler, Power Doppler, PW; Tissue Harmonic Imaging; TDI optionalShow all
Imaging technology
Mindray eWave platform; second-generation Single Crystal technology
Exam presets
Cardiac, Abdomen, Lung, TCI, EM, FAST, EM AAA, Bladder, OB/GYN, VascularShow all
AutoEF
Optional cardiac-mode measurement that draws the trace automatically
Smart Bladder
Measurement-menu function; part of the TE Air in K220051
iClear
Setting that enhances the image profile
Living Technology™
Software-based upgrades; performance updates are part of the standard warranty
Revision and software
TE Air iOS app version 4.2; earliest listed version 1.0, dated November 17, 2022
Environment
Operating environment
0 °C to 35 °C; 20% to 85% relative humidity, no condensation; 700 to 1060 hPa
Storage and transport
-20 °C to 45 °C
Compliance
FDA 510(k)
K250024 (2025-05-30), K240906 (2024-07-23), K231685 (2023-10-26), K220051 (2022-03-04)Show all
Electrical safety classification
Internally powered equipment; Type-BF applied part; continuous operation

What to check when buying a used Mindray TE Air

  1. Reject a cracked, swollen or deformed probe

    Mindray treats peeling, rifts, bumps, cracks or liquid spill on the probe as the end of its service life.

    How: Look over the probe head for cracks or swelling and the body for any deformation or damage.

  2. Have a biomed run the patient leakage test

    A probe that fails the test must be taken out of use and handled by Mindray's service engineers.

    How: After checking for liquid residue and any scorch or smoke smell, the biomed immerses the probe's acoustic head in sodium chloride solution and measures leakage with a safety analyzer.

  3. Inspect the charging port, cable and adapter

    Mindray does not supply the power adapter, so the one offered with the probe has to meet the probe's input rating.

    How: Look for damage at the probe's charging port, wrinkles, cracks or aging on the cable, and cracks or swelling on the adapter, then check that its plug seats firmly.

  4. Confirm the built-in battery charges and runs scans

    The battery sits inside the probe, and only Mindray-trained technical staff may install or remove it.

    How: Charge it with the probe switched on, scan with the probe unplugged, and if runtime seems short, read the battery self-check result.

Common questions about the Mindray TE Air

Is the Mindray TE Air FDA-cleared?

Yes. It was first cleared under 510(k) K220051 on March 4, 2022, as the TE Air Diagnostic Ultrasound System, a Class II device under 21 CFR 892.1550, for abdominal, pediatric, thoracic and pleural, adult and pediatric cardiac, neonatal and adult cephalic, and urology exams.

Three later clearances followed. K231685 (October 26, 2023) added Android support and moved the iOS requirement from iOS 12 to iOS 13; K240906 (July 23, 2024) added small organ, musculoskeletal and peripheral vessel uses, the z3P, z3Ps, z3Pt, e3P, e3Ps, e3Pt, a3Px, i5M, z5M, z5Ms, z5Mt, e5M, e5Ms, e5Mt and a5Mx transducers, and software options; K250024 (May 30, 2025) added AirSight, Mindray's customized Android platform, as a host option.

What is the difference between the TE Air i3P and the TE Air e5M?

The TE Air i3P is a phased-array probe with 64 elements, a 2.0–4.0 MHz bandwidth and 11 exam presets, for cardiac, abdominal, thoracic and pleural, urology and cephalic exams.

The TE Air e5M, announced on November 19, 2025, switches between linear and convex settings and has 22 exam presets, for cardiac, abdominal, thoracic and pleural, OB/GYN, vascular, urology, MSK, nerve and small-organ exams.

Does the Mindray TE Air need a subscription?

No. There is no subscription plan: everything is paid at purchase, and app updates arrive through the app store.

What warranty does Mindray give on the TE Air?

Bought directly from Mindray in North America, the wireless transducer has a 3-year standard warranty covering normal wear and failure, and accessories such as the USB charging cable and the Air Capsule have 1 year.

The phone or tablet that runs the app is not covered, and software performance updates through Living Technology are part of the standard warranty.